Smart Positioner with Control Valve Calibration Procedure

1-   Single Acting Control Valve with a SMART Positioner. (Direct Action and Reverse Action Actuator)

ProcedureGeneral Procedure for all Power stations and refineries.
Title of JobMaintenance Check & Calibration
ManpowerInstrument Technician Using PPE (Personal Protective Equipment)
Safety DocumentMaintenance Work Permits. & SCC (Safety Clearance Certificate) If Required
Tools/        SpecialI &C Tool Kit + any special tools (if Required)Test Equipment
Tools
Test Equipment:Note: Select test equipment according to the Test loop.
1-   4~20mADC source.
2-   Multimeter (AVO).
3-   Multifunction precision Calibrator
       
4-   HART communicator.   
5-   Air Filter Regulator (0~2bar)  
6-   Standard Pressure gauge (0~2Bar)               
Stores & MaterialsCleaning spray, brush and cloth 
Job Description
ProcessStepsDuring  Maintenance
Isolation1We  have  2  types  of    Single  acting  (Diaphragm   
&operated) control valve with a SMART Positioner   
Removal    1-  Direct Action Control Valve 2-  Reverse Action Control Valve           
   Note:(1)  SMART  positioner  is  one  touch  calibration. (2)SMART Positioner works  as 4 in 1 (It eliminates I/P converter, Pn. Positioner, Pn. position transmitter & P/I converter)           Fig. 1  
 
 2Valve should be on manual position from the auto manual station form the control room (Unit should be shut-down or annual maintenance).
 3Control valve should be isolate by isolating inlet and out let valve of the process line by operation department.
 4Carry out the control valve and accessories external cleaning, using a brush  and  approved  cleaning  spray to  remove  contamination  or  solid particles.
5Carry out the SMART positioner external cleaning, using a brush and approved cleaning spray to remove contamination or solid particles.
6Inspect control valve actuator and accessories external physical damage, general appearance & fitness (Check whether cover seals of accessories are intact).
7Inspect SMART positioner external physical damage; general appearance & fitness (Check whether cover seals are intact).
8Open the cover of SMART positioner and disconnect input and output wires carefully with core identification and insulate the all wires with insulation tape.
Preparation9Write all the details of control valve: tag no., service, and unit no. in the calibration sheet.
10Air leak test of actuator should be perform to confirm there should be no air  leak  from  anywhere  of  the  actuator.  (Leak  test  should  be  done according to the Air Leak Test Procedure Ref.: SCP-AALT-24).
11Stroke adjustment shall be checked. Re-adjust the stroke if required. (When actuator is fully close then mechanical valve should be also fully closed and there should be no passing in this situation.)
12Open the back cover of SMART positioner to clean the nozzle flapper as shown in fig. 2 Fig.2
13Clean the nozzle flapper by taking a paper with thickness of 0.5mm and move in between nozzle and flapper as shown in fig. 3 After cleaning fix back the cover.

Fig. 3
14Set up the test equipment with multifunction  calibrator  as shown in fig. 4      

Fig. 4
15Another   Set   up   the   test   equipment   can   be   used   with   HART Communicator as per the diagram shown in the fig. 5.

Note: We can also configure and calibrate SMART Positioner by this communicator.
Calibration16Insert 4mADC from the mA source or multifunction calibrator. The valve should not move from 0% Position. Check on the Mechanical indicating plate should show 0% Position.
17SMART positioner will transmit 4~20mADC signal to control room for position indicator (This will work like position transmitter) Check the position on the multifunction calibrator should be 0.0%, means it is 4.00 mA on mA meter output of positioner.
Note:  Mean time it  can  show  us  stroke in  mm  on  the multifunction calibrator if this option is available in multifunction calibrator.
18Now insert 8mA, 12mA, 16mA & 20mA accordingly from the mA source or multifunction calibrator. Let the valve should travel to 25%, 50%, 75%
& 100% Position accordingly. Check current output from the SMART
positioner to control room. It should be 8.0mA, 12.00mA, 16.00mA and
20.00mA accordingly.
Note: View on the mechanical indicating plate should also be showing all above values accordingly.
19Check  the  repeatability in  ascending and  descending order from  0  to
100% and from 100% to 0% accordingly.
20If calibration values are correct and error is acceptable limit.  Record these input & output values in calibration sheet in as calibration columns or before calibration column and go to step 33.
21If input values not matching with output position values means calibration is required.
Note: SMART positioner calibration is very easy. It is called one touch calibration. (Automatic calibration will be done by SMART positioner)
22Set 18mADC on the multifunction calibrator as shown in step 14 in fig.4.
23Take a screw driver (-) and turn the screw full right  turn  (Upward  as  shown  on  the  plate.). Hold for 3 seconds as shown in this fig.6.
When valve move to closed position leave the screw. Screw will come back its original position.

Fig. 6
Note: (1) Automatic calibration will start.
(2) Now only view the automatic calibration. No need to touch.
 24In 1st step valve will move to 0% Position. Now SMART positioner will start to adjust C/V at 0%. Also output for position transmitter will be adjusted  at  0%  Position.  Wait  till  it  stabilizes  the  all  values  at  0% position.
 25In 2nd step valve will move to 100%. SMART positioner will adjust the valve 100% position of C/V. Wait till it stabilizes the all values of 100% Position.
 26In 3rd  step valve will move again to 0% position to confirm the 0% position.
 
 27In 4th step valve will move to 50% Position. This means calibration is over.
    
 28Now confirm the 5 point calibration. Insert the  0%, 25%, 50%, 75% and 100% input current and view the valve position on the indicating plate and also view the position on the multi-function calibrator or output 4~20 mADC on multimeter.
 29We   can   also   confirm   5   points   calibration   
 Manual position of Auto / Manual screw form Auto to Manual position as shown in fig. 7.
             
    
 30Now operate the air filter regulator manually for 0%=0.2bar,  25%=0.4bar,  50%=0.6  bar, 75%=0.8 bar and 100%=1.0 bar. as shown in fig.8.

Note: indicating  plate  also  view  output  current 4~20 mADC on multimeter.
                          
                       
   
 31In case of over traveling movent of C/V more than 100%. Adjsut the strok by terning UP and Down Screw of SMART Poitioner as shown in fig. 6. Note: Up and down screw should move up and down but not hold on up and down position.
 32If calibration values are correct and error is acceptable limit.  Record these input & output values in calibration sheet in after calibration columns.
 33The control valve calibration input / output values should be inspected by Inspector and Quality Inspector for witness and record these values to sign the certificate after completion of job.
Completion34Once the test is completed, remove the test equipment and clean the tested device
 35Reconnect the input / output wires with core identification to SMART positioner.
36Reconnect the instrument fittings, tubing without bending or damaging and ensure that connector is not cross fitted that can damage threading. (If removed)
37Open  the  main  air  supply  valve  to  the  regulators  and  adjust  output pressure accordingly.
38Check  for  leakage  of  air  during  commissioning  and  retighten  the instrument fittings.
39Complete the check and calibration sheet and handover to the concerned I&C Engineer for inspection and signature.

Leave a Comment

error: Content is Protected.